Radio-Counter
32 P, 123 I, 125 I, 131 I, 99 Tc, 55 Cr, 59 Fe, 137 Cs, 18 F, 11 C, 15 O, etc.

Shell, Jr. 3000 

Click here for Brochure


Radio-Single Sample Counter for all isotopes - β , g

High energy Gamma & Positron detection

 

Applications

n Liquid sample counting

n Reference sample counting

n RIA-counting

n Solid sample counting

n Freeze-dried sample counting

n Homogenized sample counting

n Saturation thickness counting

n Waste counting

Background subtraction
Enhances the detectability of very low activities.

DPM calculation
With the entry of individual values of efficiency can be performed automatically.

High count rates
For high count rates up to 200,000 cps the system is equipped with a dead time correction.

Print out
All parameters as well as results can be obtained via RS 232C port.

Features

DISPLAY: 6 digit LED display with dimmer control providing a range of 0 - 999999 counts

LINEARITY: Reading within plus or minus 2% of true value

TIMER: Adjustment from 0 - 999 minutes with selectable divisions of X0.1, and X1

METER 1: 0 - 500,000 CPM range

METER 2: 0 - 500 CPM; 0 - 2.5 kV; BAT TEST

MULTIPLIERS: X1, X10, X100, X1000

LINEARITY: Reading within plus or minus 1% of true value with detector connected

RESPONSE: FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading

ZERO: Push button to zero meter

VOLTAGE: Adjustable from 200 - 2500 volts (will support 60 megohm scintillation loads)

THRESHOLD: Voltage sensitive; adjustable from 1.0 - 10.0

WINDOW: Adjustable from 0 - 10.0 above threshold setting (can be turned on or off)

Discriminator: Adjustable from 2 - 100 mV at a threshold setting of 1.0

RS-232: 9 pin connector allowing for printer, or software interface and Analog output

Detectors:  1"x1" NaI crystal, 2"x2" Na) crystal or BGO
Plastic vial, Stainless steel body. Can also be used for HPLC with flowcell.