Shell,
Jr. 3000
Radio-Single Sample Counter for all
isotopes - β , g
High energy Gamma & Positron detection
Applications
n Liquid sample counting
n Reference sample counting
n RIA-counting
n Solid sample counting
n Freeze-dried sample counting
n Homogenized sample counting
n Saturation thickness counting
n Waste counting
Background subtraction
Enhances the detectability of very low activities.
DPM
calculation
With the entry of individual values of efficiency can be
performed automatically.
High
count rates
For high count rates up to 200,000 cps the system is
equipped with a dead time correction.
Print
out
All parameters as well as results can be obtained via RS
232C port.
Features
DISPLAY: 6 digit LED display with dimmer control providing a range of 0 - 999999 countsLINEARITY: Reading within plus or minus 2% of true value
TIMER: Adjustment from 0 - 999 minutes with selectable divisions of X0.1, and X1
METER 1: 0 - 500,000 CPM range
METER 2: 0 - 500 CPM; 0 - 2.5 kV; BAT TEST
MULTIPLIERS: X1, X10, X100, X1000
LINEARITY: Reading within plus or minus 1% of true value with detector connected
RESPONSE: FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading
ZERO: Push button to zero meter
VOLTAGE: Adjustable from 200 - 2500 volts (will support 60 megohm scintillation loads)
THRESHOLD: Voltage sensitive; adjustable from 1.0 - 10.0
WINDOW: Adjustable from 0 - 10.0 above threshold setting (can be turned on or off)
Discriminator: Adjustable from 2 - 100 mV at a threshold setting of 1.0
RS-232: 9 pin connector allowing for printer, or software interface and Analog output
Detectors:
1"x1" NaI crystal, 2"x2" Na) crystal or
BGO
Plastic vial, Stainless steel body. Can
also be used for HPLC with flowcell.